5 Key features
5.5 Logging
Honeywell 2024
94
reached, the oldest records will be overwritten with the newer ones. The consequence of a
particular allocation setting is reflected in the number of Days worth of data as well as the
Number of records that can be written before overwriting will occur.
Note: A log’s interval setting will impact Days, but not Numbers.
l
The Enable checkboxes may be used to suspend data collection for one or more logs. A log’s
existing records will not be affected when the log is disabled. The fact that a log is disabled will be
reflected in the Log Configuration window by the presence of a red ‘x’ at the top of that log’s item
tree. In addition, the area for that log in the bar graphic will be grayed and marked as ‘Disabled’
l
Finally, click Send to Instrument to push the log configuration to the EC350 Device. A few seconds
will be required for the transfer. A window should appear indicating ‘Successfully Configured’.