Input Calibration 
5.4  Input 1 or 2 Set Up Wiring  
5.4.1 Thermocouple Inputs Using an Ice Bath 
Refer to Figure 5-2 and wire the controller according to the procedure given in Table 5-3. 
Table 5-3  Set Up Wiring Procedure for Thermocouple Inputs  
Using an Ice Bath 
Step  Action 
1 
Connect the copper leads to the calibrator. 
2 
Connect a length of thermocouple extension wire to the end of each copper lead and 
insert the junction points into the ice bath. 
3 
Connect the thermocouple extension wires to the terminals for Input #1.  
See Figure 5-2. 
 
 
+
 
 
Millivolt
 
Source
 
Ice Bath
 
Copper Leads
 
Thermocouple
 
Extension Wire
 
XXXX 
26 + 
27 - 
25 
23 + 
24 - 
22 
Input 1
Input 2
 
Figure 5-2  Wiring Connections for Thermocouple Inputs Using an Ice Bath 
 
136  UDC3200 Universal Digital Controller Product Manual  12/04