EasyManua.ls Logo

Honeywell UDC3200 - Table 6-4 Auxiliary Output Calibration Procedure

Honeywell UDC3200
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Calibration
Procedure
The procedure for calibrating the auxiliary output is listed in Table 6-4.
Make sure “LOCK” in the Tuning Set Up group is set to “NONE” (see Subsection 3.4).
Table 6-4 Auxiliary Output Calibration Procedure
Step Operation Press Result
1
Enter Calibration
Mode
SetupSetup
until you see
Upper Display = CALIB
Lower Display = AUX OUT
2
Calibrate 0 %
Function
FunctionFunction
You will see:
Upper Display = A Value
Lower Display = ZERO VAL
or
until the desired 0 % output is read on the milliammeter.
Normally, this will be the setting that produces 4 mA.
3
Calibrate 100 %
Function
FunctionFunction
To store the 0 % value you will see:
Upper Display = A Value
Lower Display = SPAN VAL
or
until the desired 100 % output is read on the milliammeter.
Normally, this will be the setting that produces 20 mA.
4
Function
FunctionFunction
The controller stores the span value.
Exit the
Calibration Mode
To exit the calibration mode.
Lower
Display
Lower
Display
Lower
Display
12/04 UDC3200 Universal Digital Controller Product Manual 147

Table of Contents

Related product manuals