9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Bit Condition Operation
0 Align RF
skipped
During an alignment, the calibration of the RF section (frontend) of the
instrument was not performed. This can be caused by an interfering
user signal present at the RF Input
See "Align Now" on page 490, "Align Now All" on page 492
Filter Registers
–
"Questionable Calibration Skipped Condition" on page 1140
–
"Questionable Calibration Skipped Enable" on page 1141
–
"Questionable Calibration Skipped Event Query" on page 1141
–
"Questionable Calibration Skipped Negative Transition" on page 1141
–
"Questionable Calibration Skipped Positive Transition" on page 1142
Questionable Calibration Skipped Condition
Returns the decimal value of the sum of the bits in the Questionable Calibration
Skipped Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote
Command
:STATus:QUEStionable:CALibration:SKIPped:CONDition?
Example
:STAT:QUES:CAL:SKIP:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
IQ Analyzer Mode User's &Programmer's Reference 1140