6 General Programming
228 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
Table 34: Control
Bit
Width
Meaning
Data
Idle
31
1
Data/command selection
0: Data
1: Command (type of command is selected by command code)
X
X
30
1
End Marker Sequence
X
X
29
1
End Marker Scenario
X
X
28
1
Init Marker Sequence
X
X
27:25
3
Reserved
X
X
24
1
Marker Enable
X
N/A
23:20
4
Advancement Mode Sequence
0: Auto
1: Conditional
2: Repeat
3: Single
4 15: Reserved
X
X
19:16
4
Advancement Mode Segment
0: Auto
1: Conditional
2: Repeat
3: Single
4 15: Reserved
X
N/A
11:0
12
Reserved
X
X
Table 35: Sequence loop count
Bit
Width
Meaning
Data
Idle
31:0
32
Number of sequence iterations (1..4G-1), only applicable in the first
entry of a sequence
X
X
Table 36: Segment loop count
Bit
Width
Meaning
Data
Idle
31:0
32
Number of segment iterations (1..4G-1)
X
N/A