4-8 2G0947 - 31JAN08
Preventive Maintenance
19. Clean and inspect:
PROCESSOR COVER ROLLERS
HEAT SHIELD SUPPORT BARS
DENSITOMETER GUIDE
DENSITOMETER GUIDE ROLLERS
20. Vacuum:
TOP COVER
FRONT PANEL
entire PROCESSOR AY
21. Remove the used piece of film from the top of the OPTICS AY.
NOTE: Be careful not to spill loose debris resting on top of the film. Discard
the film and debris.
22. Install:
PROCESSOR DRUM
RIGID DUCT
IMPORTANT:
Install a new FELT PAD AY every PM (20,000 cycles) and a new FILM
DIVERTER AY every other PM (40,000 cycles).
PROCESSOR COVER
ROLLERS
DENSITOMETER GUIDE
ROLLERS
2 HEAT SHIELD
SUPPORT BARS