5-10 2G0947 - 31JAN08
Diagnostics
Severity Levels
The following table describes the severity levels associated with condition codes
for the IMAGER.
20
21
22
23
24
25
26
27
28
29
30 - 35
36
MCS:
- General MCS System
- Upper FILM SUPPLY
- N/A For The 5800 IMAGER
- Lower FILM SUPPLY
- N/A For The 5800 IMAGER
- RF Tag
- FILM TRANSPORT
- OPTICS
- PROCESSOR
- DENSITOMETER
- N/A For The 5800 IMAGER
- LOCAL PANEL User Interface
20-xxx
21-xxx
23-xxx
25-xxx
26-xxx
27-xxx
28-xxx
29-xxx
36-xxx
Table 5-5: Subsystem IDs (Cont’d)
Subsys ID Subsystem/Component Name Error ID
Table 5-6: Severity Levels
Severity Description Recovery
0-Fatal A failure occurred and the IMAGER will
have to be restarted.
The IMAGER must be restarted.
1-Alert A failure occurred, but it should be
temporary and correctable by the user.
The IMAGER should be able to continue after
the problem is corrected. If the failure is FILM
SUPPLY specific, the IMAGER should be able
to continue operating using the other FILM
SUPPLY.
2-Warning An event occurred that should notify the
user.
No recovery needed.
3-Notify An event occurred, but was automatically
recovered.
A notice is written to the log. The IMAGER
should be able to continue automatically.
4-Auxiliary Status information. Does not impact IMAGER operation.