2G0947 - 31JAN08 5-11
Diagnostics
Condition Code: 01-000
Description
A general, non-specific error has occurred in the Configuration and Control
Subsystem.
System Impact
Unknown.
User Action
None
FE Action
None, this is a log notification only.
Condition Code: 01-001
Description
When a subsystem or component generates a non-recoverable, Configuration
and Control will attempt to shutdown and restart it. This error means that
Configuration and Control has initiated more than the configured number of
allowed restarts.
System Impact
The system will continue to run normally, unless the subsystem or component in
question cannot run any longer.
User Action
None
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
None General Error in Configuration
and Control Subsystem.
NORMAL 3-Notify
Web Portal Message Log Message DICOM Message Severity
None Maximum Number of Restarts
Exceeded for XXXX. [where
XXXX is the subsystem name]
NORMAL 2-Warning