5-22 2G0947 - 31JAN08
Diagnostics
Condition Code: 05-002
Description
Study reclamation has been initiated because either the high-water mark has
been reached, or the scheduled time for study reclamation has occurred.
System Impact
Completed studies will be systematically removed from the system until the
low-water mark is reached.
User Action
None
FE Action
None, this is a log notification only.
Condition Code: 05-003
Description
The Storage Recovery Subsystem could not reach the low-water mark while
running the storage reclamation routines.
System Impact
None. The MIM Core should continue to run normally, although this warning
may continue to be logged periodically.
User Action
None
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
None High water-mark reached.
Scheduled study reclamation
time reached.
NORMAL 3-Notify
Web Portal Message Log Message DICOM Message Severity
None Low water-mark could not be
reached.
NORMAL 3-Notify