2G0947 - 31JAN08 5-23
Diagnostics
Condition Code: 05-004
Description
Study reclamation has been completed because the low-water mark was
successfully reached.
System Impact
None.
User Action
None
FE Action
None, this is a log notification only.
Condition Code: 05-005
Description
The Study Reclamation Subsystem has received a shutdown request.
System Impact
None. Most likely, the entire system is shutting down. Also, it may be that a
configuration information change that requires the subsystem to be restarted.
User Action
None
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
None Low-water-mark reached. NORMAL 3-Notify
Web Portal Message Log Message DICOM Message Severity
None Study reclamation is shutting
down.
NORMAL 3-Notify