2G0947 - 31JAN08 5-83
Diagnostics
Condition Code: 26-544
Description
This error occurs if the trailing edge is not detected at the FILM AT
DENSITOMETER SENSOR (S11) as the film exits the DENSITOMETER. The
error is cleared when the TOP COVER is opened then closed. When this error
occurs, the PROCESSOR stops rotating and the HEATER is turned off. All
films in the printer will be marked bad and reprinted automatically.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. All other films will be reprinted.
User Action
1. Remove any films jammed in the IMAGER film exit (Area 3).
2. Open the TOP COVER and remove any films in Area 2.
3. Close the TOP COVER.
FE Action
1. Clear the jam (see “Film Jam - Error Code 26-544” on page 5-113).
Web Portal Message Log Message DICOM Message Severity
Film Jam in Area 3 Film Jam at Densitometer: Film
at Densitometer sensor (S11)
failed to detect film trailing
edge.
FILM JAM 1-Alert