5-86 2G0947 - 31JAN08
Diagnostics
Condition Code: 27-123
Description
This error occurs when the polygon speed measured by the OPTICS MODULE
MICROCONTROLLER is out of range. The error clears only when power is
cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Condition Code: 27-601
Description
This error occurs when the power monitor offset measured by the optics
calibration and diagnostics is outside the predefined valid range. The error clears
when a calibration is requested or power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Optics: Internal Hardware
Failure
Polygon fails to rotate at desired
speed
ELEC DOWN 0-Fatal
Web Portal Message Log Message DICOM Message Severity
Optics: Calibration Failed Optics Calibration Failed.
Power Monitor Offset outside
valid range
ELEC DOWN 0-Fatal