2G0947 - 31JAN08 5-85
Diagnostics
Condition Code: 26-933
Description
This error occurs when the FILM PATH MICROCONTROLLER tries to
communicate with one of the STEP MOTOR DRIVER (SMD)
MICROCONTROLLERS but receives no response to 5 attempts. The error
clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Condition Code: 26-934
Description
This notification occurs when the FILM PATH MICROCONTROLLER tries to
communicate with one of the STEP MOTOR DRIVER (SMD)
MICROCONTROLLERS but receives no response. (Note: Error 26-933 is
generated when this problem is detected on the 5th successive failed
communication attempt.)
System Impact
None - the system will function normally.
User Action
None
FE Action
Web Portal Message Log Message DICOM Message Severity
Film Transport: Internal
Communications Failure
Film Path Microcontroller failed
to communicate with an SMD
Microcontroller after multiple
attempts
ELEC DOWN 0-Fatal
Web Portal Message Log Message DICOM Message Severity
None Film Path Microcontroller failed
to communicate with an SMD
Microcontroller
NORMAL 3-Notify