2G0947 - 31JAN08 5-89
Diagnostics
Condition Code: 27-611
Description
The laser power gain is measured using a voltage on the LASER DRIVER. This
error occurs when the LASER DRIVER voltage cannot be adjusted to within
limits by the optics calibration after 20 successive attempts.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Condition Code: 27-613
Description
The system measures the end of scan jitter caused by variations in the
POLYGON. The system uses these measurements to correct for variations and
improve end of scan jitter. This error will occur if these measurements are
outside of a range of values. If this error occurs, jitter correction is disabled. The
diagnostic disables correction by setting all facet corrections to a predefined
value. These measurements fine tune the jitter seen at the end of scan and
typically the presence or absence of jitter correction is difficult to observe.
System Impact
None - the system will function normally.
User Action
None
FE Action
Web Portal Message Log Message DICOM Message Severity
Optics: Internal Hardware
Failure
Optics calibration laser power
gain adjustment failed
ELEC DOWN 0-Fatal
Web Portal Message Log Message DICOM Message Severity
None Spinner jitter correction outside
valid range (correction disabled)
NORMAL 3-Notify