2G0947 - 31JAN08 5-115
Diagnostics
LED6* LIFT_HOME LOWER - When lit, lower FILM SUPPLY PICKUP BAR is at home (up).
LED7* FILM_SAVER UPPER - When lit, upper FILM SAVER is not present.
LED8* FILM_SAVER LOWER - When lit, lower FILM SAVER is not present.
LED9, LED10 FILM_OUT UPPER, FILM_OUT LOWER - not used, never lit.
LED11* FILM_CONTACT UPPER - When lit, the upper PICKUP is not touching film in the upper
FILM SUPPLY.
LED12* FILM_CONTACT LOWER - When lit, the lower PICKUP is not touching film in the
lower FILM SUPPLY.
LED13 LASER_ILOCK - When lit, this LED indicates LASER power is enabled. It does not mean
the LASER is energized. LASER power is enabled only when:
The 3 INTERLOCK SWITCHES are closed and,
The SERVICE SWITCH is in the Normal position.
LED14* FILM_SEP UPPER - When lit, there is no film separation (the SUCTION CUPS are not
lifting a sheet of film) from the upper FILM SUPPLY.
LED15* FILM_SEP LOWER - When lit, there is no film separation (the SUCTION CUPS are not
lifting a sheet of film) from the lower FILM SUPPLY.
LED16 DOOR_ILOCK - When lit, the 24 V motor drive voltage (Hazard 24 V) is enabled. This
voltage is enabled when: .
The 3 INTERLOCK SWITCHES are closed or,
The SERVICE SWITCH is in the Service position.
LED17, LED18 TRAY_LOCK UPPER, TRAY_LOCK LOWER - not used, never lit.
LED19* FILM_PATH2 - not used, always lit.
LED20* NIPOC_LIMIT, UPPER - Indicates that the NIP ROLLER is moving from open to closed
or from closed to open.
LED21 NIPOC_LIMIT, LOWER - not used, never lit.
LED22 FILM_PATH3 - not used, never lit.
LED23* CASS_IN UPPER - When lit, there is no FILM TRAY in the upper FILM SUPPLY.
LED24* CASS_IN LOWER - When lit, there is no FILM TRAY in the lower FILM SUPPLY.
LED25 FILM_PATH4 - not used, never lit.
LED26, LED27 BAG UPPER, BAG LOWER - not used, never lit.
HAZARD +24V Test Point for Hazard 24 volts (MOTOR supply voltage). This voltage is enabled when:
The 3 INTERLOCK SWITCHES are closed or,
The SERVICE SWITCH is in the Service position.
Table 5-7: FPM LEDs and Test Points (Cont’d)