5-116 2G0947 - 31JAN08
Diagnostics
* Indicates FILM SUPPLY SENSOR LEDs - - - Each of these LEDs is controlled by one of the FILM SUPPLY
SENSORS. LED ON = Related SENSOR blocked.
+24V TEST Test Point for Digital 24 volts.
+12V TEST Test Point for Digital 12 volts.
+V5 TEST Test Point for Digital 5 volts.
+V3_3 TEST Test Point for Digital 3.3 volts.
GND1, GND2,
GND3, GND4
Test Points for Logic Ground.
JP2 (Jumper pins) Jumper across both pins overrides the MICROPROCESSOR SENSOR control and enables
all FILM SUPPLY SENSORS.
S1 RESET SWITCH - Pressing this SWITCH causes a Checksum and Memory test to run for
the software on the FPM.
S2 DEFAULTS SWITCH.
Table 5-7: FPM LEDs and Test Points (Cont’d)
Item Description