Chapter 8: T-Uptake/FTI-Measurements LB 9507
8-4
T1
READY
MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the button located
directly below. The following display appears:
T2
PROTOCOL
MASTER CURVE CREATE CHANGE
The <MASTER CURVE> option can only be used when a
“quantitative protocol” without a standard series has been created
and saved (see section 7.2).
Select <CHANGE> to open an existing protocol by entering its
protocol number. You can then edit it.
To create a new protocol, select <CREATE>. You will see this:
T3
09: PROTOCOL NAME FTI-V1
The next free number is displayed for the new protocol (in our
example “09:”). Enter a name for this protocol using the
alphanumeric keyboard. You should use a name which specifies
the protocol type and other important features, so that it is easy to
identify later. For example “FTI-V1”.
Confirm your entry with <enter>.
T4a
09: PROTOCOL TYPE
RAW DATA CUTOFF -OTHERS-
Select <OTHERS>.
T4b
09: PROTOCOL TYPE
QUANT. TUP/FTI -OTHERS-
Select <TUP/FTI >. Now you have to enter the parameters which
are the same for all protocol types: