LB 9507 Chapter 10: Kinetic Measurements
10-11
10.2.3 Performing Repeated Measurements
To perform a repeated measurement, first you have to enter a
protocol as described in chapter 10.2.2.
K1
READY
MEASURE PROTOCOL -OTHERS-
Select <MEASURE>.
K21
SELECT TYPE OF MEASUREMENT ?
PROTOCOLS DIRECT ENTRY RATEMETER
Select <PROTOCOLS> to open a stored protocol.
K22
ENTER PROTOCOL NO. 6
PRINT LIST
Enter the number of the protocol required and press <enter>.
You can print out the list of stored protocols including number and
name by pressing the button below <PRINT LIST>.
K23
06: PROTOCOL Repeated?
YES NO
You are prompted to confirm that this is the required protocol. If
you select <NO> the program returns to display K22. <YES>
leads to:
K24
COMMENT
You can enter a comment on the following measurement using
the alphanumeric keyboard. Confirm it with <enter>. If you do not
want to enter a comment, just press <enter>.
K25
SAMPLE 1 TUBE # 1

Table of Contents