EasyManua.ls Logo

LUMAT LB 9507 - INSTRUMENT PARAMETER> Menu; TURN ONCE> Function; Service Functions

LUMAT LB 9507
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4: Structure and Operartion of Software LB 9507
4-8
4.2.5 <INSTRUMENT PARAMETER> Menu
Enter and verify the instrument parameters, e.g. the volume of the
injectors used or the calibration factors (see section 3.8). Fig. 4-6
shows you how to select this menu.
4.2.6 <TURN ONCE> function
When you select this function, the measurement chamber rotates
through 180°. This allows you to move a sample tube into the
measurement position or out of the measurement position while
not actually measuring (useful when cleaning).
READY
MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
Fig. 4-7: Selecting the <TURN ONCE> function
4.2.7 Service functions
The program contains the following service functions; these are
only accessible to Berthold Technologies’ customer service
engineers.
<INJECT. TEST> Injector test.
<PM TEST> Tests the photomultiplier.
<ROTOR TEST> Tests the measurement chamber rotor.
<MONITOR> Loads new operating software.

Table of Contents