12.2 RS 232 Data Output LB 9507 in Protocol-Specific Measuring Mode
Start of measurement series
? <Prot. no.> (CR) (LF) 3 digits
<Gen. note> (CR) (LF) 0 - 15 digits
Measured standards
<RLU> (CR) (LF) for all:
.NSB
.TOTAL
. STANDARD/CAL.
. 12 digits
Use of stored standard values
O <RLU> (CR) (LF) For all mean values from
the previously measured
and stored standard
values preceded by "O“.
Edited standard or calibrator mean values
E <RLU> (CR) (LF) For all edited values,
12 characters,
E <RLU> (CR) (LF) with preceding "E“
Control samples
* <Ctr. ID> (CR) (LF) (0 ... 11 digits)
<RLU> (CR) (LF) (for each replicate)
% <Conc> (CR) (LF) (for each replicate)
: <Conc> (CR) (LF) (mean value)
Patient samples
<Pat.No.> <RLU> (CR) (LF) (for each replicate,
pat. no. 4 digits)
% <Conc> (CR) (LF) (for each replicate)
: <Conc> (CR) (LF) (mean value)