Chapterl 1: System Description LB 9507
1-10
1.5 Data Management and Software update
The Lumat LB 9507 Software is based on two parts:
❑ The kernel provides all the basic or elementary functions to
control all the electronic functions, also the download
manager and serial number.
❑ The User interface which controls the user functions and all
individual parameters eg. individual measurement times and
measurement methods, reagent timing, data management
for results and set up of evaluation procedures.
This arrangement allows you to load updates and upgrades with
the minimum of expense.
If a power failure occurs, the measured data from the last 30
measurement runs remain stored in a battery-backed memory for
at least 3 months.
The measurement protocols are stored in the instrument’s current
and battery independent internal memory, and can only be lost if
the CPU develops a fault or during a program is update.
A RS232 interface allows the kernel or operating software to be
loaded from a PC to Lumat.
The Update software will be provided only from Berthold
Technologies directly or from by Berthold Technologies authorised
service organisations.