LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-3
7.2 Setting Up Quantitative Protocols
Start by entering the parameters for the measurement procedure
and the evaluation strategy in a measurement protocol in the
<PROTOCOL> menu, and then save them. Later, you can run a
measurement simply by calling up one of these stored protocols in
the <MEASURE> menu and pressing the <START> button.
You can save up to 40 different measurement protocols which
include all information such as injector control, definition of
replicates, measuring time etc., as well as information on data
evaluation. Each new protocol is given a number which is
dependent upon the time the protocol was set up but not on the
protocol type.
The differences between the three measurement strategies are
evident in the protocol structure.
For measurements with a complete standard series, you will need
the entire protocol that is described on the following pages.
For measurements with a master curve the protocol finishes at
display L30 (see Fig. 7-1). The master curve is then entered in the
submenu <MASTER CURVE> and assigned to the respective
protocol.