Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-14
7.4 Entering a Master Curve
Protocols without complete standard series (terminate the
protocol at display L31) require a so-called master curve, the data
for which is supplied by the reagent kit manufacturer.
Enter this data as follows:
L1
READY 9 FEB 00 14:27
MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the respective
button. Then follows:
L2
PROTOCOL
MASTER CURVE CREATE CHANGE
Select <MASTER CURVE>:
L2a
ENTER PROTOCOL NO. 11
PRINT LIST
If you like you can print the list of stored protocols by selecting
<PRINT LIST>. Enter the required protocol number and press
<enter>.
Enter the desired protocol number, which must include a
quantitative protocol without full standardization, and press
<enter>.
If these requirements are not met, an error message is displayed
which you can acknowledge by pressing <exit>.
If the protocol number is accepted by Lumat the following display
appears:
L3a
PROTOCOL # 11 "LIA"
YES NO
Select <YES>. Press the button below <NO> if you want to select
another protocol number.