LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-15
L41
08: LOT ID
Enter the lot number or name using a maximum of 5 characters
and press <enter>. (Delete incorrect characters by pressing
<delete>).
L42
08: NO. OF CURVE POINTS (4-10) 4
LAST DISPLAY
Enter the number of curve points corresponding to the master
curve and press <enter>.
You can only enter RLU values in continuously rising or falling
order. Lumat does not accept illegal entries and alerts you
accordingly.
L43
08: MASTER CURVE RLU 1
LAST DISPLAY
Enter the RLU values for the 1st curve point.
L44
08: MASTER CURVE CONC. 1
LAST DISPLAY
Enter the concentration value for the 1st curve point
corresponding to the units in the respective measurement
protocol.
L45
08: MASTER CURVE RLU 2
LAST DISPLAY
Enter the RLU values for the 2nd curve point.