Chapter 3: Getting Started LB 9507
3-2
3.4 Switching on Lumat
Please watch the display:
1. After the power is switched on, the following messages
should appear:
ROTATING TO INITIALISE
When this message appears, the tube transport rotor will
perform a whole revolution. The ”ready” message then
appears with the main menu:
READY <DATE> <TIME>
MEASURE PROTOCOL -OTHERS-
If not, the microprocessor has detected an error (see Chapter
14.1).
2. If "MEMORY ERASED" is displayed together with the
"READY" message after the power is switched on, the stored
data has been lost due to a memory defect. This will also be
indicated by a 3 second beep. You can work with the
instrument after re-entering all configuration and instrument
parameters.