LB 9507 Chapter 3: Getting Started
3-9
3.6.2 Changing the Instrument Parameters
The instrument parameters of the LB 9507 are edited as follows:
READY <DATE> <TIME>
MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
SYST.CONFIG. INSTR.PARAM. -OTHERS-
ENTER CALIBRATION FACTOR FOR RED RLU 1.00
PRINT PARAM.
ENTER CALIBRATION FACTOR FOR BLACK RLU 1.00
PRINT PARAM.
These values must match the RLU value of the test protocol. It
refers to the inherent sensitivity of the photomultiplier. Confirm the
value by pressing <enter>.
If a printout of the instrument parameters is required, press the
button beneath <PRINT PARAM.>.
ENTER VOLUME INJ. 1 300
PRINT PARAM. LAST DISPLAY
Enter the volume of injector 1 in µL. This display appears only for
instruments with at least one injector. If you enter “0”, injector 1 is
disabled for all measurements.
ENTER VOLUME INJ. 2 [µl] 200
PRINT PARAM. LAST DISPLAY
Enter the volume of injector 2 in µL. This display appears only for
instruments with 2 injectors. If you enter “0”, injector 2 is disabled.
Next, you are asked: