LB 9507 Chapter 11: Operator Functions
11-3
11.1.1 Loading reagents
This function is used to fill the injector system. Three injections
with the largest possible volumes will follow.
W1
REAG. FUNC.:
PRIME REFRESH -OTHERS-
Select <PRIME>.
W2
LOAD REAG.
INJ1 INJ2
Select the injector to be filled.
W3
INSERT TUBE
Place an empty tube for the outflowing liquid. After pressing
<START> the tube will move into the measurement position and
the loading function will be run.
W4
INJECTION 1 1…3 Fold
Three injections will be performed.
W5
TURN TO THE NEXT HOLE…
REMOVE TUBE…
Remove the tube. The injector selection is displayed again (see
W2). Press the <exit> button to stop the event.