LB 9507 Chapter 3: Getting Started
3-1
3. Getting Started
3.1 Instrument Setup
The place where the instrument is set up should have the
following characteristics:
- free of dust
- no strong magnetic fields
- no direct exposure to sunlight
- ambient temperature +15 - +30°C
- relative humidity 10 - 90 %, no condensation
- power supply as specified on the type label
- distance of rear panel from the wall at least 10 cm to provide
the necessary cooling
- operation area above the instrument at least 20 cm
- firm, smooth surface leaving the air filter inlet open
3.2 Power Supply
A selection of the line voltage is not necessary. The
multi range power supply will detect the line voltage
automatically , only make sure that the correct line fuse
is fitted.
1 Amps slow blow for 230 V
2 Amps slow blow for 115 V
at a frequency of 47 - 62 Hz.
Use only the cable supplied with the instrument to
connect it to the mains!
3.3 Checking the Paper Supply
Check if the paper supply is sufficient; for changing paper see
section 13.1.