LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-17
7.5 Performing Quantitative Measurements
7.5.1 Measurement with a Complete Standard Curve
L1
READY <DATE> <TIME>
MEASURE PROTOCOL -OTHERS-
Select <MEASURE>.
L50
SELECT TYPE OF MEASUREMENT
PROTOCOLS DIRECT ENTRY RATEMETER
Select <PROTOCOLS> to open a stored protocol.
L51
ENTER PROTOCOL NO. 8
PRINT LIST
You can print out the list of stored protocols including number and
name by pressing the button below <PRINT LIST>.
Enter the number of the desired protocol (including complete
standard series) and press <enter>.
L52
06: PROTOCOL LIA-T1 O.K. ?
YES NO
You are prompted to confirm that this is the protocol you want. If
you select <NO> the program returns to display L51. <YES> leads
to: