Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-16
L46
08: MASTER CURVE CONC. 2
LAST DISPLAY
Enter the concentration of the 2nd curve point, and so on for all
the curve points defined in L 45.
An error message will be displayed if the entered value is smaller
or equal to the previous value; this prevents incorrect entries.
L47
08: LOW CAL. CONC.
LAST DISPLAY
Enter the concentration of the low calibrator (to readjust the
master curve during measurement).
L48
08: HIGH CAL. CONC.
LAST DISPLAY
Enter the concentration of the high calibrator.
L49
INPUT ALL CORRECT ? YES
YES NO LAST DISPLAY
<YES>: the entered parameters are stored and the program returns
to the main menu (L1).
<NO>: the program returns to the 1st display of the protocol (L41).
Entries and changes are stored only if you terminate the routine via
L49 with <YES>.