Chapter 10: Kinetic Measurements LB 9507
10-14
10.3 Ratemeter
The Lumat LB 9507 allows continuous kinetic measurements that
are not based on stored protocols and provide a high degree of
flexibility.
Raw data are measured continuously in this ratemeter mode, and
injections can be performed manually at any time. At the same
time - in accordance with the defined measuring intervals - results
are printed, including a record of the injection times.
K1
READY
MEASURE PROTOCOL -OTHERS-
Select <MEASURE>.
K32
SELECT TYPE OF MEASUREMENT
PROTOCOLS DIRECT ENTRY RATEMETER
Select <RATEMETER>.
K33
RATEMETER
ENTER MEASURING TIME [s] 5.0
Enter the desired measuring time in seconds (= measuring
intervals). The measured value (RLU/s) determined in each
interval is shown on the display and can be printed optionally. The
data can also be continuously sent to a PC through the serial port
(I/O port), where they can be evaluated.
The measurement is started by confirming the measuring time
with <enter>. The following display then appears:
K34
RATEMETER [RLU/s] 55
TURN NEXT INJECTING PRINT ON
Here you have the following choices: