Chapter 11: Operator Functions LB 9507
11-2
11.1 <REAGENT> Function
In the <REAGENT> menu the following reagent specific functions
are available:
<PRIME> Loading the injector system with reagents with a pre-defined
number of injetion cycles
<REFRESH> Refreshing the reagent in the injector system. This function is
needed for example when the luminometer has not been used for
a long period of time.
<WASH> The injector system can be cleaned with this function
<MANUAL UNLOAD> Pumps back the reagents from the injector system into the
reagent bottles
OPEARATOR FUNCTIONS
REAGENT PERF. TEST -OTHERS-
REAG. FUNC.:
PRIME REFRESH -OTHERS-
OPER. FUNC.: DATA EVALUATION:
WASH MANUAL UNLOAD -OTHERS-