Chapter 11: Operator Functions LB 9507
11-4
11.1.2 Refreshing reagents
This function is used to refresh the reagents in the injector
system. If the luminometer has not been used for a long period of
time, the injector volume may have changed due to evaporation or
changes in pressure. Refreshing is performed to refill the reagent
correctly.
W1
REAGENT:
PRIME REFRESH -OTHERS-
Select <REFRESH>.
W2
REFRESH
INJ1 INJ2
Select the injector for the refreshing function.
W3
INSERT TUBE
Place an empty tube for the outflowing liquid. After pressing
<START> the tube will move into the measurement position and
the refreshing function is run:
W4
INJECTION 1…
One injection is performed.
W5
TURN TO NEXT...
REMOVE TUBE…
Remove the tube. The injector selection is displayed again (see
W2). Press the <exit> button to stop the event.