LB 9507 Chapter 10: Kinetic Measurements
10-15
<PRINT ON>
This function prints all measurement results. The display toggles
to <PRINT OFF>.
<TURN NEXT>
The next tube is moved into the measuring chamber.
<INJECTOR 1> or <INJECTOR 2>
(only when 1 injector selected)
If only one injector is active, pressing the button below it at any
time injects into the tube in the measuring chamber.
<INJECTING> (only when 2 injectors are active)
This function produces the following display:
K35
RATEMETER [RLU/s] 55
TURN NEXT INJECTION 1 INJECTION 2
Here you can actuate the injectors at any time. The tube transport
function remains on the display.
For safety reasons an injection can be performed only after
the measurement chamber with an empty tube in it has
turned once.
Select <exit> to return to display K34.
Each injection is printed in the right order in the printout.
Note:
If the measuring intervals are short, the printer may not be able to
keep up with the speed of the measurement procedure, so the
measurements and printing will be delayed. The “Injection...”
message on the printer may also often come much later than the
actual injection. You should therefore use the current
measurement result on the display as a basis for deciding when to
perform an injection.
Interrupt the kinetic measurement by selecting <exit> when K34 is
displayed.