Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-4
L1
READY
MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the respective
button. Then follows:
L2
PROTOCOL
MASTER CURVE CREATE CHANGE
The <MASTER CURVE> option can be used only when a
“quantitative protocol” without a standard series has been created
and stored. If you press the button below <CHANGE> you can
open an existing protocol by entering its protocol number, and edit
it. To create a new protocol, press the button below <CREATE>.
The following display is shown:
L3
08: PROTOCOL NAME LIA-T1
The next free number is displayed for the new protocol (in this
example “08:”). Enter a name for this protocol using the
alphanumeric keyboard. You should use a name which specifies
the protocol type and other important features, so that it is easy to
identify later. For example “LIA-T1”.
Confirm your entry with <enter>.
L4a
08: PROTOCOL TYPE
RAW DATA CUTOFF -OTHERS-
Select <OTHERS>.
L4b
08: PROTOCOL TYPE
QUANT. TUP/FTI -OTHERS-
Select <QUANT>. Now you have to enter the parameters which
are the same for all protocol types: