LB 9507 Chapter 9: Dual-Luciferase
TM
Reporter Gene Assay
9-17
D24
COMMENT
You can enter a comment on the following measurement using
the alphanumeric keyboard. Confirm it with <enter>. If you do not
want to enter a comment, just press <enter>.
D28
REMOVE TUBE...
Remove the tube from the sample holder if there is one. This
display appears if there is a tube in the sample holder. So, if a
tube already is in the correct position before you start
measurement, you will have to lift the tube briefly. The display will
show:
D28a
SAMPLE 1 TUBE # 1
INSERT TUBE...
Load the tube containing the first sample as defined in the
protocol.
D29
SAMPLE 1 TUBE # 1
START
Press the button beneath the <START> function or the <start>
button to the right of the rotor. Both buttons do the same thing.
The tube will move back into the measurement chamber, and the
programmed measurement procedure will begin:
D30
BACKGROUND MEASUREMENT...
Background measurement is performed if you specified it in the
protocol.