LB 9507 Chapter 11: Operator Functions
11-15
G4
EVALUATE RUN 28 WITH PROT 07 ?
YES NO
You are prompted to confirm whether to evaluate the selected
measurement with the same protocol that was used for the
original measurement or manual entry.
Normally, you will select <YES> here, but in special cases you
can also select (the ‘dangerous’ option) <NO>, which lets you
evaluate the (raw) data with a protocol other than the one used for
the original measurement or manual entry. This can be useful if a
measurement or evaluation was mistakenly done with a wrong
protocol, or a false master curve was used. If you select <NO>
you must enter another protocol number, and display G4 again
prompts you to confirm. In both cases you reach display G5.
G5
COMMENT OLD COMMENT
The originally entered remarks are retrieved from the rerun
memory and used as the default value, which you can edit or
confirm. Always conclude with <enter>. You then reach G6.
G10
FETCH LOW CAL. REPL. ... TUBE ...
The low calibrator(s) of the original measurement are loaded and
printed here.
G11
FETCH HIGH CAL. REPL. # TUBE #
The low calibrator(s) of the original measurement are loaded and
printed here.