11 Troubleshooting
CLEAN THE INSTRUMENT
The exterior casing of the instrument may be cleaned using a clean, lint-free cloth dampened with
isopropyl alcohol (IPA), a mild detergent, or a 3% hydrogen peroxide solution. Do not use any type
of abrasive cleaner. It is not necessary to remove knobs, screws, etc. while cleaning.
n
Use only a mild detergent in water to clean safety shields. The use of isopropyl alco-
hol can damage the shield surface.
n
Do not allow liquid to penetrate the casing of the analyzer. Doing so could result in
damage to the unit.
n
Use only a mild detergent in water to clean safety shields. The use of isopropyl
alcohol can damage the shield surface.
CLEAN THE GAS DELIVERY TUBES
If using a FlowPrep, gas delivery tubes should be wiped with a clean, lint-free cloth after each use
to remove any particles of sample that may have adhered to the tube. If a gas delivery tube
becomes clogged or damaged, it should be replaced.
1. Remove the gas delivery tube from the flexible tubing.
2. Attach a new gas delivery tube by pressing the flexible tubing over the gas delivery tube.
Ensure that the flexible tubing completely covers the ferrule at the end of the gas delivery
tube.
LUBRICATE THE ELEVATOR DRIVE ASSEMBLY
The elevator screw is lubricated before it leaves the factory and should not require lubricating. If
the elevator starts to vibrate or becomes noisy when traveling, contact a Micromeritics Service
Representative for disposition.
Should lubrication become necessary, apply a light coat of Teflon Magnalube-G grease
[Micromeritics part number 004-16163-00] to the elevator screw. Do not grease the guide rods.