EasyManuals Logo

Micromeritics TRISTAR II PLUS User Manual

Micromeritics TRISTAR II PLUS
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #105 background imageLoading...
Page #105 background image
PERFORM A KRYPTON ANALYSIS
Unit [n] >Krypton Analysis
In 21CFR11 environments, users are required to login to the Confirm application to
start an analysis. Once the analysis window is opened, manual control is disabled
until the analysis has completed. During analysis, pausing and resuming is allowed,
however, steps cannot be skipped.
Krypton analyses are available only if the krypton option is installed.
Data collection is done sequentially one analysis starts and completes before the
next analysis begins.
Use to perform up to three simultaneous krypton analyses. Use the following guidelines for
performing krypton analyses:
n
All analyses must use krypton gas. An error message occurs if a gas other than krypton is
specified.
n
From Psat tube must be selected as the Dosing Method on Adsorptive Properties. An error
message displays if Normal is selected as the Dosing Method. Samples cannot be
removed from or added to ports until the full set of analyses has completed.
1. Click Browse to select a sample file.
2. Click Report after analysis to automatically generate reports when the analysis is com-
plete. On the Report Settings window, select the report destination. Click OK.
3. Click Start to start the analysis. A window displays data as they are collected. A short delay
is encountered before the port status changes from the Idle state at the bottom of the
window. When analysis is complete, remove the sample tube and store (or dispose) of the
sample material as applicable.
4. Click Start to start the analysis. A window displays data as they are collected. A short delay
is encountered before the port status changes from the Idle state at the bottom of the
window. When analysis is complete, remove the sample tube and store (or dispose) of the
sample material as applicable.
Perform a Krypton Analysis
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023
6 - 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics TRISTAR II PLUS and is the answer not in the manual?

Micromeritics TRISTAR II PLUS Specifications

General IconGeneral
BrandMicromeritics
ModelTRISTAR II PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals