11 Troubleshooting
PERFORM A LEAK TEST
Unit [n] > Diagnostics > Start Diagnostic Test
A Micromeritics Service Representative may request that a leak test be performed to determine if
there is a system leak and may also require a copy of the report generated by this test.
The test provides:
n
Prompts on preparing the analyzer for the test.
n
Approximate time period of the test.
n
Prompts in which an operator response is required.
1. Select the test to run.
2. Select Report After Test and choose Preview as the destination.
3. Click Start.
4. Verify all tests have a Passed status by selecting the tabs and looking for the Passed status
for each test run.
5. Click Save As to save the test file results.
11 - 12
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023