8 Selected Report Options
BJH ADSORPTION/DESORPTION REPORT
The BJH calculation determines the mesopore volume/area distribution, which accounts for both
the change in adsorbate layer thickness and the liquid condensed in pore cores. Reports can be
generated from both adsorption and desorption data. The fields for both BJHAdsorption Report
Options and BJHDesorption Report Options are identical unless otherwise specified.
An incomplete pore distribution may be generated if a thickness curve selection is not
a good match for the sample being analyzed.
Circled selection is applicable to BJH
Adsorption only
Selections
Description
Adsorptive [button]
Displays the Adsorptive Options window. The recommended adsorpt-
ives and their values are shown. Up to eight adsorptive and
adsorbate property factor combinations may be specified.
BJHAdsorption/Desorption Report
8 - 8
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023