1 Analyzer Components for TriStar IIPlus
EQUIPMENT OPTIONS AND UPGRADES
Parts and accessories are located on the Micromeritics web page.
Option Description
Chiller Dewar
A closed loop recirculating system that utilizes a high surface area
copper coil to provide excellent heat transfer between the Dewar and
the recirculating liquids. The Chiller Dewar Quick Start Guide is
located on the Micromeritics web page.
FlowPrep
The FlowPrep applies both heat and a stream of inert gas to the
sample to remove adsorbed contaminants from the surface and
pores in preparation for analysis for up to six samples. Choose the
temperature, gas, and flow rate best suited for the sample material.
The FlowPrep is an independent unit and not controlled by the ana-
lyzer.
Krypton Option
For very low surface area analyses. The option consists of a high-
vacuum pump, and 10mmHg transducer.
Smart VacPrep
The Smart VacPrep prepares samples by heating and evacuation. It
contains six sample ports in which up to five temperatures, ramp
rates, and soak times per sample are individually controlled by the
analyzer program so that all degas information is integrated into the
sample data file for future reference. Samples can also be prepared,
started, and completed independently. There is no need to wait for
samples on other ports to finish. Front panel buttons allow a
QuickStart operation with preprogrammed conditions.
Up to three additional Smart VacPrep degassers can be connected
to one computer permitting 24 preparation ports to be used. The
Smart VacPrep is the recommended degassing unit.