T-PLOT REPORT
The t-Plot calculation allows quantitative analysis of the area and total volume ascribed to
micropores. Matrix area (the area external to micropores) is directly determined and often proves
to be a valuable way of characterizing complex mixed materials.
Selections
Description
Fitted thickness range
[textbox]
Enter the minimum and maximum thicknesses (in angstroms or nano-
meters) to include in the thickness curve. Go to Options > Units to
specify default units.
Pressures [button]
Use to select a pressure range for report calculations and points for
exclusion from calculations.
Calculation pressure range. Enter the minimum and maximum
pressures to be used in the pressure table.To exclude a point from
the calculations used to generate the report, select Exclude.
Exclude All.Select to exclude all pressure points in the table.
Include All.Select to include all pressure points in the table.
t-Plot Report
t-Plot Report
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023
8 - 51

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics TRISTAR II PLUS and is the answer not in the manual?

Micromeritics TRISTAR II PLUS Specifications

General IconGeneral
BrandMicromeritics
ModelTRISTAR II PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals