INSTRUMENT STATUS
SHOW DASHBOARD
Unit [n] >Show Dashboard
Data for the dashboard comes from the logged diagnostic data. The dashboard remains current
as the relevant diagnostic data items are updated. The gauges will be updated even if the
dashboard window is not open.
Red numbers on the dashboard require attention. To reset the dashboard numbers,
right-click on the dashboard setting, then click Reset.
Selections
Description
Analyses
completed / started
Displays N/M where N is the number of analyses that have finished
data collection and M is the number of analyses that have been star-
ted. Analyses canceled or terminated by errors before the termination
stage starts are not counted as completed.
Days until roughing-
pump service is due
Annual maintenance is recommended. The number of days until the
anniversary of the last pump maintenance is shown. The displayed
value is updated at least once per day and when the maintenance
time is reset. When the displayed value is 30 or less, the value is dis-
played in red. Red negative numbers display if maintenance is past
due.
Dashboard Gauges
Instrument Status
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023
2 - 17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics TRISTAR II PLUS and is the answer not in the manual?

Micromeritics TRISTAR II PLUS Specifications

General IconGeneral
BrandMicromeritics
ModelTRISTAR II PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals