MP-METHOD REPORT
The MP-Method Report provides pore volume distributions for microporous materials by
correlating quantity adsorbed with the thickness of the adsorbed layer as determined from a user-
selected thickness curve. Pore size can be expressed in angstroms or nanometers. Go
toOptions >Units to specify the unit.
Selections
Description
Pressures [button]
Use to select a pressure range for report calculations and points for
exclusion from calculations.
Calculation pressure range. Enter the minimum and maximum
pressures to be used in the pressure table.To exclude a point from
the calculations used to generate the report, select Exclude.
Exclude All.Select to exclude all pressure points in the table.
Include All.Select to include all pressure points in the table.
MP-Method Report
MP-Method Report
TriStar II Plus Operator Manual