2 About the Software
Selections
Description
Manifold outgas rate
Provides the qualitative indication of the outgas rate in the dosing
manifold. LED images constitute a bidirectional bar graph of the
outgas rate.
The gauge is updated when the Analysis Manifold Test is run. See
Start Diagnostic Test on page9 - 1.
n
Three green LEDs are lit if outgas rate is below 30% of outgas rate
limit.
n
At 30%, the left LED turns off.
n
At 60%, the center LED turns off.
n
At 90%, three green LED lights turn off and the center yellow
LEDturns on.
n
At 110% and above, only the red LEDturns on and attention is
required.
Nitrogen P
0
Displays statistics of the saturation pressures measured with nitrogen
gas at liquid nitrogen temperatures. The mean, two-sigma, minimum,
and maximum values display. The gauge is updated when a P
0
is
logged with nitrogen as the adsorptive and a bath temperature of
77±2K.
Dashboard Gauges (continued)
2 - 18
TriStar II Plus Operator Manual