BJH Correction
[groupbox]
Select the type of correction to apply to calculations. The selected
type will display in the report header.
Faas. Good for statistical thickness curves.
Kruk-Jaroniec-Sayari. Good for reference thickness curves.
Standard. Uses original BJH models.
Cumulative Reports
[groupbox]
Larger. Use to report the total volume found in pores larger than the
current pore size.
Smaller. Use to report the total volume found in pores smaller than
the current pore size.
Pores [groupbox]
Enter the minimum and maximum diameter (radius or width) of pores
to include in the BJH reports.
Fraction of pores open at both ends. This field is not available for
the BJH Desorption Report Options window.
During adsorption calculations, the analysis program assumes that
all pores are closed at one end. Occasionally, a percentage of pores
may be open at both ends causing disagreement in the adsorption
and desorption data or in the values for total volume and total BJH
pore volume. Enter the fraction of pores open at both ends to
compensate for this error.
BJHAdsorption/Desorption Report (continued)
BJH Adsorption/Desorption Report
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023
8 - 9