A - 18 BeneVision N Series Patient Monitor Operator’s Manual
A.15.7 NIBP Specifications
Standard Meet standard of ISO 80601-2-30
Technique Oscillometry
Mode of operation Manual, Auto, STAT, Sequence
Auto mode repetition intervals 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time 5 min
Max measurement time Adult, pediatric: 180 s
Neonate: 90 s
Heart rate range 30 to 300 bpm
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 25 to 290 25 to 240 25 to 140
Diastolic: 10 to 250 10 to 200 10 to 115
Mean: 15 to 260 15 to 215 15 to 125
Accuracy Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure range
(mmHg)
Adult: 80 to 280
Pediatric: 80 to 210
Neonate: 60 to 140
Default initial cuff inflation pressure
(mmHg)
Adult: 160
Pediatric: 140
Neonate: 90
Software overpressure protection Adult: 297±3 mmHg
Pediatric: 297±3 mmHg
Neonate: 147±3 mmHg
Static pressure measurement range 0 mmHg to 300 mmHg
Static pressure measurement accuracy ±3 mmHg
PR
Measurement range 30 to300 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Alarm limit Range (mmHg)
Step (mmHg)
NIBP-S High Adult: (low limit + 5) to 285
Pediatric: (low limit + 5) to 235
Neonate: (low limit + 5) to 135
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Low 26 to (high limit - 5)
NIBP-M High Adult: (low limit + 5) to 255
Pediatric: (low limit + 5) to 210
Neonate: (low limit + 5) to 120
NIBP-M Low 16 to (high limit - 5)
NIBP-D High Adult: (low limit + 5) to 245
Pediatric: (low limit + 5) to 195
Neonate: (low limit + 5) to 110
NIBP-D Low 11 to (high limit - 5)