BeneVision N Series Patient Monitor Operator’s Manual A - 19
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in
terms of mean error and standard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
NIBP-S Extreme High Adult: (NIBP-S high limit + 5) to 290
Pediatric: (NIBP-S high limit +5) to 240
Neonate: (NIBP-S high limit +5) to 140
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Extreme Low 25 to (NIBP-S low limit - 5)
NIBP-M Extreme High Adult: (NIBP-M high limit + 5) to 260
Pediatric: (NIBP-M high limit + 5) to 215
Neonate: (NIBP-M high limit + 5) to 125
NIBP-M Extreme Low 15 to (NIBP-M low limit - 5)
NIBP-D Extreme High Adult: (NIBP-D high limit + 5) to 250
Pediatric: (NIBP-D high limit + 5) to 200
Neonate: (NIBP-D high limit + 5) to 115
NIBP-D Extreme Low 10 to (NIBP-D low limit - 5)