7-6 TM80 Telemetry Monitor Service Manual
Item Alarm Message Possible cause Solution
NIBP
NIBP Error An error occurred to the
NIBP module.
There is a problem with
communication
between the TM80 and
BP10.
Restart the BP10.
NIBP Cuff Loose The NIBP cuff is not
properly connected.
There is a leak in the
airway.
Check the patient's
condition and verify
patient category.
Replace with an
appropriate cuff and
connect it correctly.
NIBP Airway Error An airway error occurs. Check the airway.
NIBP Weak Signal The patient’s pulse is
weak or the cuff is loose.
Check the patient’s
condition and change
the cuff application
site. If the error
persists, replace the
cuff.
NIBP Overrange The measured NIBP
value is not within the
specified range.
Contact your service
personnel.